MODEL 6300VISIBLE RANGESPECTROPHOTOMETEROPERATING MANUAL630 026
7The complimentary colour method of selecting wavelength may not be applicable in all situations, perhapsbecause the solution has no distinct colour,
8Concentration ModeThis mode is selected by moving the cursor to the CONC menu option using the LEFT or RIGHT arrowkeys. The primary display will show
3.3 GOOD PRACTICE GUIDELINES 1. For optimum performance a calibration routine should be carried out at the beginning and end ofevery sample batch.2. T
SECTION 4MAINTENANCE4.1 GENERALThe 6300 has been designed to give optimum performance with minimum maintenance. It is only necessaryto keep the extern
Fig. 4.2.1 Lamp Fitting6. Close the lamp access panel and re-tighten the thumbscrew.NOTE:It is essential that only the specified replacement lamp shou
SECTION 6INTERFACINGSerial InterfaceThe Model 6300 has a bi-directional RS232 interface set to:1200 baud7 data bitsodd parity1 stop bitThe 25 way D co
ASCII Fxxxx.x<CR> Sets the concentration factor to xxxx.x. For example: F1000<CR> will set thefactor to 1000.Note <CR> is an ASCII c
EC Declaration of ConformityJenway Model 6300 Spectrophotometer complies with the following European Standards:EN 50081-1:1992 Electromagnetic compati
6300/REV B/12-96SAFETYPlease read this information carefully prior to installing or using this equipment.1. The unit described in this manual is desig
6300/REV B/12-966300 SPECTROPHOTOMETEROPERATING MANUALCONTENTSSECTION 1 INTRODUCTIONInstrument Description 1.1Instrument Specification 1.2SECTION 2 IN
SECTION 1INTRODUCTION1.1 INSTRUMENT DESCRIPTIONThe Model 6300 is a microprocessor controlled Visible Range Spectrophotometer covering the wave-length
SECTION 2INSTALLATION2.1 UNPACKINGRemove the Model 6300 from the packaging and ensure the following items are present:1. Model 6300 Spectrophotometer2
2.3 DISPLAYS 1. Primary display area - Transmission, Absorbance, Concentration 2. Primary display adjust annunciator 3. Secondary display area - Wavel
2.4 CONTROLS1. used to adjust values on the selected display2. used to move horizontally between menu options3. used to select the displayed menu opti
2.5 INPUTS/OUTPUTSFig. 2.5.1 Rear Panel Layout 1. LAMP ACCESS PANEL This panel allows the user to gain access to thetungsten halogen lamp when replace
SECTION 3OPERATION3.1 INITIAL SET-UPNOTE: If the unit is in continuous use it is recommended that it should be left switchedon to obtain maximum lamp